Squidstat: Difference between revisions
Started section on EIS |
Expanded EIS section, and added Nyquist plot subsection |
||
| Line 33: | Line 33: | ||
The Squidstat Plus is a single channel combined potentiostat galvanostat with FRA and EIS capabilities. It is capable of applying ±10V of potential to a sample or ±1A of current. EIS measurements can be made within the 10 µHz - 2 MHz frequency range. It has 5 alligator clip cables: 2 are lead used to apply potential and current, one attaches to the working electrode and the other to counter electrode; 2 are sense leads used for measuring potential and current at the working and counter electrodes; the fifth is a reference electrode, used to determine what potential will be assigned to 0 V. | The Squidstat Plus is a single channel combined potentiostat galvanostat with FRA and EIS capabilities. It is capable of applying ±10V of potential to a sample or ±1A of current. EIS measurements can be made within the 10 µHz - 2 MHz frequency range. It has 5 alligator clip cables: 2 are lead used to apply potential and current, one attaches to the working electrode and the other to counter electrode; 2 are sense leads used for measuring potential and current at the working and counter electrodes; the fifth is a reference electrode, used to determine what potential will be assigned to 0 V. | ||
== Galvanostat == | |||
A galvanostat applies a potential to an electrical circuit or cell and measures a response in the current. | A galvanostat applies a potential to an electrical circuit or cell and measures a response in the current. | ||
== Potentiostat == | |||
A potentiostat applies a current to an electrical cell or circuit and measures the resulting response in potential. | A potentiostat applies a current to an electrical cell or circuit and measures the resulting response in potential. | ||
== Electrochemical Impedance Spectrophotometry (EIS) == | |||
Electrochemical impedance spectrophotometry (EIS) is the application of a sinusoidal potential or current wave to an electrochemical cell in order to determine the impedance of the circuit. The application of a potential wave causes a corresponding current wave in the cell, and vis versa with the application of a current wave. The response wave is always separated from the applied wave by some phase shift φ. Measuring this phase shift allows the impedance of the cell to be | Electrochemical impedance spectrophotometry (EIS) is the application of a sinusoidal potential or current wave to an electrochemical cell in order to determine the impedance of the circuit. The application of a potential wave causes a corresponding current wave in the cell, and vis versa with the application of a current wave. The response wave is always separated from the applied wave by some phase shift φ. Measuring this phase shift allows the impedance of the cell to be characterized. | ||
Impedance is essentially the generalized form of resistance. Ohms Law, | Impedance is essentially the generalized form of resistance. Ohms Law, <math> \text{R} = \frac{\text{V}}{\text{I}} </math>. Ohms Law is accurate only when applied to an ideal resistor, which simplifies the relationship<ref>https://www.gamry.com/application-notes/EIS/basics-of-electrochemical-impedance-spectroscopy/</ref>. In any other circuit or electrical device, Ohms Law becomes less precise, necessitating the use of impedance. Impedance, like resistance, measures the resistance of a circuit to electrical flow. It differs from resistance in that it is dependent on frequency, and the current and potential waves may not necessarily be in phase with each other. When Ohms law is used to calculate resistance, independence of frequency and in phase waves are assumed. | ||
Impedance has 2 components, the phase shift (<math> \phi </math>) in the wave caused by the applied signal, and the amplitude (<math> Z_0 </math>) of the resulting wave. Impedance is represented by the letter '''Z''' and consists of a real and imaginary part. | |||
=== Nyquist Plots === | |||
Nyquist plots are often used to represent impedance. These are generated through EIS testing across a range of frequencies. The imaginary component of impedance is plotted on the y axis and the real component on the x axis. A parametric curve drawn from the frequencies tested is plotted. Nyquist plots are popular because they allow the components of the system to be easily identified, and it is easy to see key data points. However, the phase shift is not shown on Nyquist plots. | |||
Revision as of 16:46, 20 June 2025
| Admiral Instruments Squidstat Plus [SOP] | |
|---|---|
| Voltage Scan Range | ±10 V |
| Measured Voltage Accuracy | 0.1% of measured, 1 mV min |
| Measured Voltage Resolution | 0.003% of range, 15 μV min |
| Current Range | ± 1 A |
| Measured Current Accuracy | 0.1% of range, 100 pA min |
| Measured Current Resolution | 0.003% of range, 3 pA min |
| EIS Frequency Range | 10 µHz - 2 MHz |
| Time Resolution | 10 ns |
The Squidstat Plus is a single channel combined potentiostat galvanostat with FRA and EIS capabilities. It is capable of applying ±10V of potential to a sample or ±1A of current. EIS measurements can be made within the 10 µHz - 2 MHz frequency range. It has 5 alligator clip cables: 2 are lead used to apply potential and current, one attaches to the working electrode and the other to counter electrode; 2 are sense leads used for measuring potential and current at the working and counter electrodes; the fifth is a reference electrode, used to determine what potential will be assigned to 0 V.
Galvanostat
A galvanostat applies a potential to an electrical circuit or cell and measures a response in the current.
Potentiostat
A potentiostat applies a current to an electrical cell or circuit and measures the resulting response in potential.
Electrochemical Impedance Spectrophotometry (EIS)
Electrochemical impedance spectrophotometry (EIS) is the application of a sinusoidal potential or current wave to an electrochemical cell in order to determine the impedance of the circuit. The application of a potential wave causes a corresponding current wave in the cell, and vis versa with the application of a current wave. The response wave is always separated from the applied wave by some phase shift φ. Measuring this phase shift allows the impedance of the cell to be characterized.
Impedance is essentially the generalized form of resistance. Ohms Law, . Ohms Law is accurate only when applied to an ideal resistor, which simplifies the relationship[1]. In any other circuit or electrical device, Ohms Law becomes less precise, necessitating the use of impedance. Impedance, like resistance, measures the resistance of a circuit to electrical flow. It differs from resistance in that it is dependent on frequency, and the current and potential waves may not necessarily be in phase with each other. When Ohms law is used to calculate resistance, independence of frequency and in phase waves are assumed.
Impedance has 2 components, the phase shift () in the wave caused by the applied signal, and the amplitude () of the resulting wave. Impedance is represented by the letter Z and consists of a real and imaginary part.
Nyquist Plots
Nyquist plots are often used to represent impedance. These are generated through EIS testing across a range of frequencies. The imaginary component of impedance is plotted on the y axis and the real component on the x axis. A parametric curve drawn from the frequencies tested is plotted. Nyquist plots are popular because they allow the components of the system to be easily identified, and it is easy to see key data points. However, the phase shift is not shown on Nyquist plots.