X-ray Diffraction

Revision as of 10:11, 27 May 2021 by Rzhou95 (talk | contribs)
Jump to navigation Jump to search
Equinox XRD
Range: 0 to 110
Sample Type: Powdered samples

Overview

An X-ray diffractometer (XRD) detects crystalline phases in a sample by measuring the diffraction pattern of X-rays at varying angles. X-rays are used because the interplanar spacing for the vast majority of crystalline materials falls in the range of X-ray wavelengths (0.01-10 nm). The MILL’s XRD is set up for powder X-ray diffraction: for the best results, samples should be ground into a fine powder so that the collective group of crystals has a random orientation.

Operation

This tool is not available for direct usage by users. Anyone who wishes to use or be trained on the XRD must undergo the general lab safety training AND the Georgia Tech Office of Radiological Safety (ORS) X-Ray Safety Training before reporting to the XRD Technical Officer.

Otherwise, please visit Love 176 to fill out an XRD request slip and leave your sample in the XRD sample exchange box by the door. In order for your request to be processed, please email the technical officer listed on the XRD request slip to notify them.

Theory

Constructive interference of scattered x-rays in a BCC structure. The phase shifts differ by a factor of half the interplanar spacing distance.

XRDs use an X-ray source to generate a beam of X-rays that are reflected off crystalline planes in the sample material. A detector, which typically rotates around the sample, captures a set of reflected X-rays corresponding to different planar orientations. When each plane of atoms reflects the incident beam, the reflected wave has a specific sinusoidal offset, or phase shift. The reflected X-rays from all planes form a diffraction pattern.

At certain angles, planes may reflect waves with phase shifts separated by the wavelength of the incident X-ray. These reflected X-rays align and recombine into a more intense X-ray, a phenomenon known as constructive interference. Otherwise, the electromagnetic waves instead cancel each other out, undergoing destructive interference. We can measure the diffraction intensity at each angle to determine the orientation of crystals in the sample. Because crystalline materials typically contain equally separated planes, we can also use the intensity data to calculate the separation between two planes of atoms. This allows us to calculate the lattice parameter, a, for virtually any crystalline structure.

Bragg's Law

Powder X-ray diffraction data from polycrystalline NaCl; detector range from 0 to 60 degrees.

Given a constant interplanar spacing d, angle of reflection θ, X-ray wavelength λ, and any positive integer n, we can express Bragg's Law, the fundamental equation of diffraction. As seen in Figure 3, the sine of the angle of reflection represents the number of wavelengths in the corresponding phase shift over the interplanar atomic spacing.

Or more commonly: 2dsin(θ) = nλ

For a family of planes corresponding to the Miller indices {h k l}, denote the normal vector to that plane as:

N = <h k l> |N| = sqrt(k^2 + k^2 + l^2)

Note that in a cubic crystal:

d = a / |N|

For a given crystal, (λ/(2a)) is constant. Therefore:

sin(θ) = |N|(λ/(2a)) sin(θ1)/sin(θ2) = |N1| / |N2|

This relation between reflection angle and the Miller indices allows us to determine crystal structures from an XRD plot.

Data

If the angle of reflection is theta, the angle between the incident and reflected waves is 2θ. In the plot, we represent the range of angles as 2θ, on the x-axis, while detection intensity is measured on the y-axis.

For a specific family of planes, there are distinct diffraction patterns—there are two important rules which allow us to use peak data to determine if a material is BCC or FCC. For BCC structures, the sum of the Miller indices, h + k + l, must be even for diffraction to occur on the {h k l} family of planes. For the FCC structure, {h k l} must all have the same parity (all even or all odd).